AcuTie II

Product Overview

The Acute Innovations AcuTie II Sternum Closure System is a revolutionary design which encompasses the simplicity of standard wire cerclage while providing increased compression, stability in multiple planes, lateral protection, and straightforward installation.

Indications for Use

The Acute Innovations AcuTie II Sternal Fixation System is intended for use in the stabilization and fixation of fractures of the anterior chest wall, including sternal fixation following sternotomy and sternal reconstructive surgical procedures.

Greater Stability

Image

Resists motion in all three planes – lateral, rostral-caudal, anterior-posterior.

AcuTie II plates and wire cerclage were installed on foam blocks and loaded in all three planes until 2 mm of displacement was created.

AcuTie II outperformed wire significantly in all three planes.1

Image
Four cleats located in line with the force, and a crimp that maintains cerclage tension, resist lateral motion.
Image
The rigidly connected cleats penetrate into the sternum to resist rostral-caudal motion.
Image
The plate acts as a solid strut to span the osteotomy and resist anterior-posterior forces.
Image
A loop of wire provides compression and deep sternal stabilization that can only be achieved by encircling the sternum.

Lateral Protection

Directional cleats assist with distributing the load and reduces stress on lateral edges.

Directional cleats penetrate through soft tissue and into the bone for stability, like a screw, while distributing the load and reducing stresses on the lateral edges of the sternum.

Like barbs, they allow medial movement of the sternum halves during reduction and gripping during distraction.

Image

Twisted wire and AcuTie II plates were installed on simulated foam sterna with 20 pcf (cortical) exteriors and 5 pcf (cancellous) interiors.

The sternal halves were pulled laterally and the force to tear through the bulk of the sample was measured. AcuTie II could resist 73% more force than wire alone.1

Image
Image

A finite element analysis shows how load is distributed via the cleats from the lateral sternal edges to the anterior table of the sternum.

Testing shows a 73% increase in the force needed to tear through when compared to wire alone.1

Image

Easy Installation

  • Plate installs in less than 90 seconds.1
  • Quick, repeatable installation.
  • Confident re-entry without specialized instrumentation.
Image

Cyclic Performance

The AcuTie II plate experienced approximately 82% less motion than twisted wire.1

The plate’s crimp feature eliminates fatigue and breakage caused by twisted and bending wire.  At coughing forces, a twist will typically break within 20,000 cycles.  An AcuTie II crimp will survive more than 200,000 cycles.1

The crimp, combined with the other system features, results in significantly improved cyclic performance.

AcuTie II plates and twisted wire were installed on 30 pcf foam blocks and loaded from 0 to 200N for 10,000 cycles.  60% of twisted wires broke prior to reaching 10,000 cycles.  The average AcuTie II plate sample experienced approximately 82% less motion than twisted wire.1

Image
Image

Cost-Effective

Image

Other plating options can cost $1,480 to $8,508 per case depending on the system, number of implants used, and the additional OR time needed for installation.2,3,4,5

An AcuTie II plate can be installed in less than 90 seconds using standard sternotomy wire and minimal instrumentation.1

Locking plate systems can take more than 15 minutes to install.4 In addition to longer anesthesia time, and at a hospital cost of approximately $25/min, excluding surgeon time,5 this can translate to significant added cost.

High cost may make some options prohibitive for routine use, even when limited to patients at high risk for DSWI.

AcuTie II Sternum Closure System Surgical Lab and Technique Overview

This video demonstrates the installation of the AcuTie II Sternum Closure System.

Citations

  1. Data on file at ACUTE Innovations.
  2. Huh J Bakaeen F, Chu D, Wall MJ. Transverse sternal plating in secondary sternal reconstruction. J Thorac Cardiovasc Surg. 2008 Dec 1;126(6):1476-80.
  3. Levin LS, Miller AS, Gajjar AH, Bremer KD, Span n J, Mialo CA, et al. An Innovative Approach for Sternal Closure. Ann Thorac Surg. 2010 Jun:89(6):1995-9.
  4. Song DH, Lohman RF, Renucci JD, Jeevanandam V, Raman J. Primary sternal plating in high-risk patients prevents mediastinitis. Eur J Cardiothorac Surg. 2004 Aug 1;26(2):367–72.
  5. Stahl JE, Sandberg WS, Daily B, Wiklund R, Egan MT, Goldman JM, et al. Reorganizing patient care and workflow in the operating room: a cost-effectiveness study. Surgery. 2006 Jun;139(6):717–28.